NIST Calculations May Improve Temperature Measures for Microfluidics

September 8, 2009

(PhysOrg.com) -- If you wanted to know if your child had a fever or be certain that the roast in the oven was thoroughly cooked, you would, of course, use a thermometer that you trusted to give accurate readings at any temperature within its range. However, it isn’t that simple for researchers who need to measure temperatures in microfluidic systems—tiny, channel-lined devices used in medical diagnostics, DNA forensics and “lab-on-a-chip” chemical analyzers—as their current “thermometer” can only be precisely calibrated for one reference temperature. Now, researchers at the National Institute of Standards and Technology have proposed a mathematical solution that enables researchers to calibrate the “thermometer” for microfluidic systems so that all temperatures are covered.

Reactions taking place in microfluidic systems often require heating, meaning that users must accurately monitor temperature changes in fluid volumes ranging from a few microliters (a droplet approximately 1 millimeter in diameter) to sub-nanoliters (a droplet approximately 1/10 of millimeter in diameter). A common technique, for example, depends heavily on precise temperature cycling. Ordinary thermometers or other temperature probes are useless at such tiny dimensions, so some groups have turned to temperature-sensitive fluorescent dyes, particularly rhodamine B. The intensity of the dye’s fluorescence decreases with increasing temperature. The idea is that the dye can be used as a noninvasive way to map the range of temperatures occurring within a microfluidic system during heating and, in turn, provide a means of calibrating that system for experiments.

However, the technique currently requires the user to base all readings on the fluorescence at a single reference temperature. Previous groups have developed “calibration curves” that relate temperature to rhodmaine B fluorescent intensity based on a reference temperature of about 23 degrees Celsius (a technique first proposed by NIST researchers David Ross, Michael Gaitan and Laurie Locascio in 2001*). But it turns out that the curves are only good for that one temperature. In an upcoming paper in Analytical Chemistry**, the NIST team—Jayna J. Shah, Michael Gaitan and Jon Geist—reports that changing the reference point, such as the higher temperature when a microfluidic system is first heated, introduces errors when a dye intensity-to-temperature calculation is done using current methods.

“Our analysis shows that a simple linear correction for a 40 degrees Celsius reference temperature identified errors between minus 3 to 8 degrees Celsius for three previously published sets of calibration equations derived at approximately 23 degrees Celsius,” says lead researcher Shah.

To address the problem, the NIST team developed mathematical methods to correct for the shift experienced when the reference temperature changes. This allowed the researchers to create generalized calibration equations that can be applied to any reference temperature.

Microfluidic DNA amplification (production of numerous copies of DNA from a tiny sample) by the polymerase chain reaction (PCR) is one procedure that could benefit from the new NIST calculations, Shah says. “PCR requires a microfluidic device to be cycled through temperatures at three different zones starting around 65 degrees Celsius, so a useful dye intensity-to-temperature ratio would have to be based on that temperature and not a reference point of 23 degrees Celsius,” she explains.

More information:

* D. Ross, M. Gaitan and L.E. Locascio. Temperature measurement in microfluidic systems using a temperature-dependent fluorescent dye. Analytical Chemistry, Vol. 73, No. 17, pages 4117-4123, Sept. 1, 2001.

** J.J. Shah, M. Gaitan and J. Geist. Generalized temperature measurement equations for rhodamine B dye solution and its application to microfluidics. , Vol. 81, No. 19, Oct. 1, 2009 (published online Sept. 1, 2009).

Provided by National Institute of Standards and Technology (news : web)


print this article email this article download pdf blog this article bookmark this article     Stumble it Digg this share on Facebook retweet share on Reddit add to delicious
Rate this story - 5 /5 (1 vote)


September 8, 2009 all stories

Comments: 0

5 /5 (1 vote)
  • Stumble this up

  • Digg this

  • share this

  • hide
  • Related Stories

  • Micro Microwave Does Pinpoint Cooking for Miniaturized Labs
    created Nov 08, 2007 | popularity not rated yet | comments 0
  • Novel temperature calibration improves NIST microhotplate technology
    created Aug 11, 2009 | popularity not rated yet | comments 0
  • NIST Micro Sensor and Micro Fridge Make Cool Pair
    created Apr 15, 2008 | popularity not rated yet | comments 0
  • Low-cost microfluidics can be a sticky problem
    created May 12, 2006 | popularity not rated yet | comments 0
  • Driving water droplets uphill
    created Apr 02, 2008 | popularity not rated yet | comments 0



  • hide
  • Relevant PhysicsForums posts

  • a question in Lewis structure???
    created 1hour ago
  • Paint Technology
    created 17 hours ago
  • About pH, Metallic electrodes
    created Nov 24, 2009
  • Goddard
    created Nov 21, 2009
  • More from Physics Forums - Chemistry

Other News

Study shows flavanol antioxidant content of US chocolate and cocoa-containing products

Chemistry / Other

created 1hour ago | popularity not rated yet | comments 0

A recent study confirms that the antioxidants and other plant-based nutrients in chocolate and cocoa products are highly associated with the amount of non-fat cocoa-derived ingredients in the product. The study expands on ...


A sticky solution for identifying effective probiotics

Chemistry / Biochemistry

created 20 hours ago | popularity 5 / 5 (1) | comments 0

Scientists have crystallised a protein that may help gut bacteria bind to the gastrointestinal tract. The protein could be used by probiotic producers to identify strains that are likely to be of real benefit to people.


Sandia CR5

Machine Converts CO2 into Gasoline, Diesel, and Jet Fuel

Chemistry / Biochemistry

created Nov 23, 2009 | popularity 4.6 / 5 (30) | comments 19

(PhysOrg.com) -- Researchers at Sandia National Laboratories have built a machine that uses the sun's energy to convert carbon dioxide waste from power plants into transportation fuels such as gasoline, diesel, ...


New hydrogen-storage method discovered

New hydrogen-storage method discovered

Chemistry / Materials Science

created Nov 22, 2009 | popularity 4.3 / 5 (41) | comments 15

Scientists at the Carnegie Institution have found for the first time that high pressure can be used to make a unique hydrogen-storage material. The discovery paves the way for an entirely new way to approach ...


New on-off 'switch' triggers and reverses paralysis in animals with a beam of light

New on-off 'switch' triggers and reverses paralysis in animals with a beam of light (w/ Video)

Chemistry / Biochemistry

created Nov 18, 2009 | popularity 4.8 / 5 (5) | comments 4

In an advance with overtones of Star Trek phasers and other sci-fi ray guns, scientists in Canada are reporting development of an internal on-off "switch" that paralyzes animals when exposed to a beam of ultraviolet ...